Optical dissolved oxygen sensor
OXY-DIOS-DSP is a high quality process sensor of a concentration of oxygen dissolved in the wastewater, potable and tap water. The sensor ensures a quick measurement response and a minimum drift at the level of +/-0.2mgO2/L/year. In the typical operating conditions a calibration is required once a year, and a change of its measuring window once per two years. Due to the applied measurement method an oxygen probe does not require an application of electrolytes and exchangeable membranes. OXY-DIOS-DSP was designed for operating in the heavy environmental conditions. A sensor body is entirely made of acid-proof titanium steel. A substrate of the measuring windows is impact-resistant sapphire glass. The standard version contains a system of pneumatic cleaning. A periodical maintenance is not required. The sensor does not require an external transmitter what significantly reduces costs of its purchase and a degree of complexity while connecting it to the PLC controller. The integrated universal transmitter allows for a communication in the MODBUS/RS485, CANopen/CANBUS, USB standard or using the isolated current loop 4-20mA. The integrated USB port is for performing a configuration and calibration. No addition software is required. |
Features
- Optical measurement of the concentration of dissolved oxygen in water and wastewater
- Operating without a flow of water / wastewater
- High stability and measurement accuracy
- method of fluorescence quenching, polymer-glass matrix with platinum porphyrine dye
- trigonometric measurement of a phase - DSP signal processing, accuracy 1ns, ultra-short time of dye activation 4ms
- measurement accuracy +/-0,1mg/L range 0-5mg/L i +/-0.2mg/L from 5mg/L
- drift +/-0.2mgO2/L/year
- Reduce handling
- calibration ones a year without additional equipment
- replacement of the measuring window every three years
- built-in pneumatic system of measuring window cleaning
- High resistance to the environmental degradation
- housing from EN 1.4571 acid-prof titanium steel
- measuring window from sapphire glass
- no plastic components operating in the wastewater
- built-in moisture removal system
- Quick assembly and disassembly
- screw-thread M32x1.5 with doubled O-rings
- waterproof connector M12 - stainless steel
- quick connector PU 6/4 - stainless steel
- Built-in atmospheric pressure sensor
- Built-in universal RS485/MODBUS, CANBUS/CANopen, current loop 4-20mA, USB universal transmitter
- Power 24VDC +/-20%
- electronic overcurrent, overvoltage and undervoltage protection
- build-in an electronically attached terminator 120 Ohm
- electronic overcurrent, overvoltage and undervoltage protection
- Built-in controller of an air pump of a reaction tank
- Assembly on a tube support 35mm
- an attached acid-prof adapter G1" (to be welded in)
- possible installation directly to the pipe with an unprocessed thread G1"
- Sensor is supplied in the transport case as a set:
- sensor OXY-DIOS-DSP-V2
- support adapter G1"
- safe calibration tubes with sodium sulfite 1g 6pcs
- USB flash memory with data backup
- USB/M12 cable
- M12 plug - screw connectors
- blanking plug of a pneumatic quick connector
- syringe with silicone lubricant Loctite LB 8108 NSF 1ml
- vaccum filter 5um
- sensor OXY-DIOS-DSP-V2
- Three years of warranty for a sensor and measuring window
Specification
Parameter | Description | |
Type | Dissolved oxygen sensor in water | |
Model | OXY-DIOS-DSP 150105061 INTEGRON Poland EU | |
Measurement method | Optical – fluorescence quenching with frequency modulation excitation light, trigonometric measurement of the phase | |
Optical head | 447 / 650 nm with reference channel 617 nm, replaceable | |
Measuring interval | 1, 2, 5, 10 s | |
The activation time fluorophore during the measurement | 6 ms | |
Measuring window | Type | Porphyrin platinum closed in a polymerglass matrix |
Substrate | Sapphire glass dim. φ12.5 mm / 1 mm | |
Durability | Up to 2 year | |
Chemical resistance | Methanol, ethanol, isopropanol | |
No chemical resistance | Chloroform, benzene, toluene, xylene, acetone and other organic solvents |
|
Drift | 0.2 mg/L-O2 by a year, (20°C - water saturated with oxygen) | |
Calibration | One point | 0% - O2, aqueous solution of sodium sulfite Na2SO3 or nitrogen 4.0 |
Two point | 0% / 20.946% O2 (air - built-in barometer and humidity sensor) or air saturated with water vapour | |
Minimum flow | No need for flow of water / wastewater | |
Working temperature | 0 to 50°C | |
Storage temperature | -20 to 70°C | |
Maximum depth | 30 m | |
Measured parameters | Concentration, saturation, temperature, O2 partial pressure, air / hydrostatic pressure, humidity | |
Measuring range O2 | Partial pressure O2 | 0 to 400 mBar |
Concentration | 0 to 20 mg/L | |
Saturation | 0 to 200% | |
Measurement accuracy O2 | Range 0 to 5.0 mg/L, 20°C | +/-0.10 mg/L |
Range 5.0 to 20.0 mg/L , 20°C | +/-0.20 mg/L | |
Measurement resolution O2 | Range 0 to 1.0 mg/L | 0.001 mg/L |
Range from 1.0 mg/L | 0.01 mg/L | |
Response time |
21% O2 to 0% O2 (20°C - nitrogen), T90 < 40 s, |
|
Resistance to environmental factors | H2S, pH, K+, Na+, Mg2+, Ca2+,NH4+, Al3+, Pb2+, Cd2+,An2+, Cr, Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, CO2+, CN-, NO3-, SO42-, S2-, PO43-, Cl- | |
Barometric / hydrostatic pressure |
Accuracy | 0.3% |
Resolution | 0.1 mBar | |
Range | 200 to 2500 mbar (absolute) | |
Temperature | Sensor type | Pt1000 class A |
Accuracy | +/- 0.2°C | |
Resolution | 0.01°C | |
Range | -20 to +85°C | |
Response time | T90 < 120 s | |
Humidity |
Accuracy | 2.5% |
Resolution | 0.01% | |
Range | 0 to 100% | |
Materials | stainless steel-titanium EN 1.4571, stainless steel EN 1.4301, sapphire glass AL2O3, FPM, silicone | |
Dimensions | Diameter | 35 mm |
Length | 238 mm | |
Weight | 900 g | |
Mounting | Type | Thread M32x1.5, adapter M32-φ35 / L=80mm for welding in thin-walled pipe |
Sealing | 2 x O-ring 26x2 FPM / Viton, recommended grease Loctite 8104 |
|
Position | Minimum 30° from the vertical | |
Protection level | IP68 | |
Socket connection | Type | M12A, 8 pins, male, IP68 |
Recomended plug | Binder 99 0486 12 08 | |
Electromagnetic immunity | EN 61326 Class B | |
Window cleaning system | Pneumatic - compressed gas, rotating nozzles | |
Medium | Air / Nitrogen | |
Connection | Quick connector PA 6/4mm pipe | |
Working pressure | 0.7 bar + hydrostatic pressure | |
Maximum pressure at the outlet nozzle | 1.2 bar + hydrostatic pressure | |
Maximum pressure at the input of the system | 6 bar | |
Power | +24VDC, +/-20% | 0.1A inactive current outputs |
1.1A active current outputs | ||
Electrical strength |
max 80.0V | |
Overvoltage protection |
28.8V | |
Under-voltage protection |
18.2V | |
Overcurrent protection | 1.2A, electronic with auto restart | |
Battery RTC |
Lithium Li/FeS2 1.5V / 1.5Ah AAA, recommended Energizer L92 |
|
Control outputs | 24V/0.5A-Source, according IEC 61131-3, short circuit protection | |
MODBUS (RS485) Communication |
Mode | RTU, ASCII |
Speed (bps) | 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200, 230400, 460800, 921600 | |
Max number of devices | 64 (½ unit load) | |
Parity | None, even, odd | |
Terminator | Built-in 120 Ohm, attached electronically | |
CANOPEN (CANBUS) Communication |
Mode | SDO, PDO mapped |
Speed (kbps) | 10, 20, 50, 100, 125, 250, 500, 800, 1000 | |
Max number of devices | 112 | |
Suppresion | Common-mode filter | |
Terminator | Built-in 120 Ohm, attached electronically | |
Current loop 4-20mA | Range | 3.6 to 22mA, (4 to 20mA scalable) |
Resolution / Accuracy | 5.5uA / 0.05% | |
Supply voltage | max 28.8V | |
Measurement resistance | >250 Ohm (recomended 500 Ohm) | |
Galvanic isolation | 1kV | |
Alarm | NAMUR43 / 22mA | |
USB communication | Mode | USB 2.0 FS, virtual COM port, VT100 terminal emulation |
Supply voltage | 4.0V to 5.8V / 500mA | |
Warranty | Sensor | 3 years |
Measuring window | 2 years |
|
Expected lifetime | 8 years for work in municipal wastewater |
User manuals & Firmware
![]() |
OXY-DIOS-DSP-Instrukcja-PL-v.1.34 |
|
![]() |
OXY-DIOS-DSP-Manual-EN-v.1.34 |
|
![]() |
Changelog.txt |
|
![]() |
Firmware-OXY-DIOS-DSP-v1.31.bin |
|
![]() |
Firmware-OXY-DIOS-DSP-v1.30.bin |
Spare parts
|
|
Consumable parts
![]() |
Measuring windowModel: WSPT3 150105051
Price: 120 EUR (without VAT)
Measuring window for OXY-DIOS-DSP sensor. Backgroung: sapphire glass Al2O3. |
![]() |
Calibration tubeModel: Sodium sulfite Na2SO3 (5pcs) 150105065
Price: 5 EUR (without VAT)
Aqueous solution of sodium sulphite Na2SO3, micro tube, weight 1g. Dissolve a content in 500mL of demineralised water. |
![]() |
Silicon grease for O-rings
Model: Loctite 8104 150105068
Price: 3 EUR (without VAT)
Silicone grease for elastomers, resistant to warm and cold water, micro tube, volume 1mL. |
![]() |
Connector M12-8pin
Model: Conec 43-00130
Price: 13.50 EUR netto (without VAT)
M12-8pin plug for the sensor connector. Max cable diameter 8mm, min 4mm. |
![]() |
MODBUS/CANopen 3x2x0,22 120OHm Cable
Model: LAPP UNITRONIC BUS LD 3x2x0,22 2170205
Price: 1.9 EUR/m netto (without VAT)
Cable diameter 7.2mm. Impedance 120OHm. Temperature -40°C to +80°C. |